INSTRUMENT: GAS CHROMATOGRAPH IDENTIFIER: GN361999IN043 |
LABORATORY (Owner of the instrument or who performs the method):
Prof. Dr. W. Roether
Address :
Universität Bremen, FB 1, Institut fuer Umweltphysik
P.O.Box 330 440
28334 Bremen, Germany
Tel : 49 (0)421 218 3511 Fax : 49 (0)421 218 7018
Email : wroether@physik.uni-bremen.de
CONTACT SCIENTIST :
Dr. Birgit Klein, address as above, email: bklein@physik.uni-bremen.de
INSTRUMENT NAME : gas chromatograph (GC) with capilary column
MANUFACTURER : GC : Hewlett Packard 5890, main column : J&W Scientific, Trap filling :
Alltech, Valves : Valco Instruments
MANUFACTURING DATE : 1992
SERIAL NUMBER : GC Hewlett Packard 5890
DESCRIPTION :
The gaschromatographic system at Bremen has been custom designed to allow high precision measurements
of four CFC species (CFC-11, CFC-12, CFC-113 and CCl4). The combination of a capillary column with a
low-volume trap yields high resolution in the chromatograms and full seperation of peaks. The system is fully
automated and allows eight water samples and associated calibration measurements to be carried out un-
attented. Water samples are transferred into the system by means of flow-through containers, incorporating
glass ampoules. Generaly, measurements will be carried out immeadiately after sampling at sea, but
alternatively the ampoules can be flame-sealed, allowing sample storage for later analysis ashore. Standard
gas is used for calibration and all concentrations are reported on the SIO93 scale. The overall precision for
surface waters amounts to ± 0.6% (CFC-11, CFC-12) ± 1.5% (CFC-113) and ± 2.4% (CCl4). Detection limits
are 0.002, 0.002, 0.008 and 0.003 pmol/kg for CFC-11, CFC-12, CFC-113 and CCl4.
TECHNICAL CHARACTERISTICS :
The gas chromatograph is a Hewlett Packard 5890, the main column GS GASPRO (ID 0.32 mm, 30 m long) is
manufactored by J&W Scientific and the trap column (1mm ID, 10 cm long) is filled with Unibeads 2S (Alltech).
COMMENTS :
INSTRUMENT TYPE : Laboratory equipement
OTHER ATTACHED EQUIPEMENT (In case of complex multi sensor : Platform equipment): -
MEASURED PARAMETERS | CALIBRATION DATE | CRUISES | COMMENTS |
---|---|---|---|
|
BIBLIOGRAPHICAL REFERENCES (METHODOLOGY):
Bulsiewicz, K. , H. Rose, O. Klatt, A. Putzka and W. Roether, A capillary- column chromatographic system for efficient chlorofluorocarbon measurement in ocean waters, J. Geophys. Res., Vol. 103, C8, 15959-15970, 1998.
For any question or remarks please contact to Dr Birgit Klein